Display, Service – VEGA VEGABAR 55 4 … 20 mA User Manual
Page 33
33
6 Set up with the display and adjustment module PLICSCOM
VEGABAR 55 • 4 … 20 mA
36733-EN-130321
Basic adjustment
▶
Basic adjustment
1
Display
Diagnostics
Service
Info
Unit
1.1
Unit of measurement
bar
▼
Temperature unit
°C
▼
Position correction
1.2
Offset
P
=
0.2 mbar
0000 mbar
Min. adjustment
1.3
000.0 %
=
0.0 mbar
0.0 mbar
Max. adjustment
1.4
100.00 %
=
100.00 mbar
0.0 mbar
Damping
1.5
1 s
Linearization curve
1.6
Linear
▼
Sensor-TAG
1.7
Sensor
Display
Basic adjustment
2
▶
Display
Diagnostics
Service
Info
Displayed value
2.1
Pressure
▼
Displayed value
▼
2.1
Scaled
Display unit
2.2
Volume
▼
l
▼
Scaling
2.3
0 % = 0.0
l
100 % = 100.0
l
Backlight
2.4
Switched off
▼
Service
Basic adjustment
4
Display
Diagnostics
▶
Service
Info
Current output
4.1
Characteristic: 4-20 mA
▼
Fail.mode: < 3.6 mA
▼
Min. current: 3.8 mA
▼
max. current: 20.5 mA
▼
Simulation
4.2
Start simulation
▼
Reset
4.3
Select reset
▼
Language
4.4
Deutsch
▼
Copy sensor data
4.7
Copy sensor data?
PIN
4.8
Enable?
Application
4.9
Level
▼