VEGA VEGABAR 55 4 … 20 mA User Manual
Page 49
49
9 Supplement
VEGABAR 55 • 4 … 20 mA
36733-EN-130321
Display and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Adjustment elements
4 keys
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the sensor without cover IP 40
Material
Ʋ Housing
ABS
Ʋ Inspection window
Polyester foil
Voltage supply
Operating voltage
Ʋ Non-Ex instrument
12 … 36 V DC
Ʋ Ex-ia instrument
12 … 30 V DC
Ʋ Exd instrument
18 … 36 V DC
Operating voltage with illuminated display and adjustment module
Ʋ Non-Ex instrument
20 … 36 V DC
Ʋ Ex-ia instrument
20 … 30 V DC
Ʋ Exd instrument
20 … 36 V DC
Permissible residual ripple
Ʋ < 100 Hz
U
ss
< 1 V
Ʋ 100 Hz … 10 kHz
U
ss
< 10 mV
Load
see diagram
1000
750
500
250
12
18
16
14
20
22
24
26
28
30
32
34
36
V
3
1
2
Fig. 19: Voltage diagram
1 Voltage limit Ex-ia instrument
2 Voltage limit non-Ex/Ex-d instrument
3 Operating voltage
Electrical protective measures
Protection rating