VEGA VEGABAR 55 4 … 20 mA User Manual
Page 43
43
9 Supplement
VEGABAR 55 • 4 … 20 mA
36733-EN-130321
Ʋ Connection cable with IP 68 1 bar
version
PE, PUR
Weight approx.
0.8 … 8 kg (1.764 … 17.64 lbs), depending on process
fitting
Output variable
Output signal
4 … 20 mA
Signal resolution
1.6 µA
Failure signal output current
mA value unchanged 20.5 mA, 22 mA, < 3.6 mA (adjust-
able)
Max. output current
22 mA
Load
see load diagram under Power supply
Met NAMUR recommendation
NE 43
Dynamic behaviour output
Run-up time approx.
10 s
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 17: Sudden change of the process variable. t
T
: dead time; t
A
: rise time; t
S
: jump response time
1 Process variable
2 Output signal
Dead time
≤ 150 ms
Rise time
≤ 100 ms (10 … 90 %)
Step response time
≤ 250 ms (ti: 0 s, 10 … 90 %)
Damping (63 % of the input variable)
0 … 999 s, adjustable
Input variable
Adjustment
Adjustment range of the min./max. adjustment relating to the nominal measuring range:
Ʋ Percentage value
-10 … 110 %
Ʋ Pressure value
-20 … 120 %
Adjustment range of the zero/span adjustment relating to the nominal measuring range:
Ʋ zero
-20 … +95 %