VEGA VEGAPULS SR 68 4 … 20 mA_HART two-wire User Manual

Page 24

Advertising
background image

24

4 Mounting

VEGAPULS SR 68 • 4 … 20 mA/HART two-wire

38294-EN-130624

d

a

α

Fig. 17: Proposal for installation after orientation VEGAPULS SR 68

The angle of inclination depends on the vessel dimensions. It can be

easily checked with a suitable level or water leve on the sensor.
The following chart specifies the distance "a" between installation

position and vessel centre dependent on the measuring distance for

inclination angles of 2° … 10°.

Distance d

(m)

10°

2

0.1

0.1

0.2

0.3

0.4

4

0.1

0.3

0.4

0.6

0.7

6

0.2

0.4

0.6

0.8

1.1

8

0.3

0.6

0.8

1.1

1.4

10

0.3

0.7

1.1

1.4

1.8

15

0.5

1.0

1.6

2.1

2.6

20

0.7

1.4

2.1

2.8

3.5

25

0.9

1.7

2.6

3.5

4.4

30

1.0

2.1

3.2

4.2

5.3

Example:
In a vessel 20 m high, the installation position of the sensor is 1.4 m

from the vessel centre.
The necessary angle of inclination of 4° can be read out from this

chart.