6 mounting steps with housing and thread – VEGA VEGABAR 66 Foundation Fieldbus User Manual

Page 18

Advertising
background image

18

4 Mounting

VEGABAR 66 • Foundation Fieldbus

36740-EN-130321

3. Turn the lock fitting on the lower hexagon into the welded socket.

Use a resistant seal. Spanner width SW 41, torque max. 80 Nm.

4. Adjust the connection tube of VEGABAR 66 to the requested

height and hold it

5. Turn the upper hexagon into the lower hexagon. Spanner width

SW 41, torque max. 80 Nm. VEGABAR 66 is now temporarily hold

by the washer disc.

6. Tighten fixing screws (2) and (5) with an Allen wrench size 2.5.

Torque max. 7 Nm.

The fixing screws press lightly into the connection tube and fasten

VEGABAR 66 in this position

4.6 Mounting steps with housing and thread

Fig. 10: Housing and thread
1 Housing

2 Seal

3 Thread

Mount VEGABAR 66 with housing and thread in the following way:
1. Weld the welded socket G1½ or 1½ NPT to the vessel top
2. Insert the transmitter with connection tube or suspension cable

into the opening

3. Turn the thread with seal into the socket and tighten with SW 46

1)

Warning:

The housing must not be used to screw the instrument in! Applying

tightening force can damage internal parts of the housing.

Mount VEGABAR 66 with housing and thread in the following way:
1. Fasten the mounting bracket at the suitable height on the basin

wall

Information:

We recommend articles from the line of VEGA accessories:

Mounting bracket of stainless steel, article no. 2.21615

Counter nut of PP, article no. 2.10371

Mount into the vessel

Mounting into the basin

1)

Seal with resistant material with thread 1½ NPT.

Advertising