10 Supplement

VEGABAR 66 • Foundation Fieldbus

36740-EN-130321

Ʋ zero

-20 … +95 %

Ʋ span

-120 … +120 %

10)

Ʋ Difference between zero and span

max. 120 % of the nominal range

Recommended max. turn down

10 : 1

Nominal measuring ranges and overload capability in bar/kPa
The specifications are only an overview and refer to the measuring cell. Limitations due to the ma-

terial and version of the process fitting are possible. The specifications on the nameplate apply.

Nominal range

Overload capacity, max.

pressure

Overload capacity, min. pres-

sure

Gauge pressure
0 … +0.1 bar/0 … +10 kPa

+15 bar/+1500 kPa

-0.2 bar/-20 kPa

0 … +0.2 bar/0 … +20 kPa

+20 bar/+2000 kPa

-0.4 bar/-40 kPa

0 … +0.4 bar/0 … +40 kPa

+30 bar/+3000 kPa

-0.8 bar/-80 kPa

0 … +1 bar/0 … 1+00 kPa

+35 bar/+3500 kPa

-1 bar/-100 kPa

0 … +2.5 bar/0 … +250 kPa

+50 bar/+5000 kPa

-1 bar/-100 kPa

0 … +5 bar/0 … +500 kPa

+65 bar/+6500 kPa

-1 bar/-100 kPa

0 … +10 bar/0 … +1000 kPa

+90 bar/+9000 kPa

-1 bar/-100 kPa

0 … +25 bar/0 … +2500 kPa

+130 bar/+13000 kPa

-1 bar/-100 kPa

Absolute pressure
0 … 1 bar/0 … 100 kPa

35 bar/3500 kPa

0 bar abs.

0 … 2.5 bar/0 … 250 kPa

50 bar/5000 kPa

0 bar abs.

0 … 5 bar/0 … 500 kPa

65 bar/6500 kPa

0 bar abs.

0 … 10 bar/0 … 1000 kPa

90 bar/9000 kPa

0 bar abs.

0 … 25 bar/0 … 2500 kPa

130 bar/13000 kPa

0 bar abs.

Nominal measuring ranges and overload capability in psig
The specifications are only an overview and refer to the measuring cell. Limitations due to the ma-

terial and version of the process fitting are possible. The specifications on the nameplate apply.

Nominal range

Overload capacity, max.

pressure

Overload capacity, min. pres-

sure

Gauge pressure
0 … +1.5 psig

+200 psig

-3 psig

0 … +3 psig

+290 psig

-6 psig

0 … +6 psig

+430 psig

-12 psig

0 … +15 psig

+500 psig

-15 psig

0 … +35 psig

+700 psig

-15 psig

0 … +70 psig

+950 psig

-15 psig

0 … +150 psig

+1300 psig

-15 psig

0 … +350 psig

+1900 psig

-15 psig

10)

Values less than -1 bar cannot be set.