VEGA VEGABAR 66 Foundation Fieldbus User Manual

Page 52

Advertising
background image

52

10 Supplement

VEGABAR 66 • Foundation Fieldbus

36740-EN-130321

Ʋ Ohmic contact

Between ground terminal, process fitting and transmitter

Ʋ Connection cable between IP 68

housing and external electronics

PUR, FEP, PE

Ʋ Type label support with IP 68 version

on cable

PE hard

Weight approx.

Ʋ Basic weight

0.7 kg (1.543 lbs)

Ʋ Suspension cable

0.1 kg/m (0.07 lbs/ft)

Ʋ Connection tube

1.5 kg/m (1 lbs/ft)

Ʋ Straining clamp

0.2 kg (0.441 lbs)

Ʋ Threaded fitting

0.4 kg (0.882 lbs)

Lengths

Ʋ Connection tube

0.25 … 6 m (0.82 … 19.69 ft)

Output variable
Output

Ʋ Signal

digital output signal, Foundation Fieldbus protocol

Ʋ Physical layer

according to IEC 61158-2

Channel Numbers

Ʋ Channel 1

Primary value

Ʋ Channel 2

Secondary value 1

Ʋ Channel 3

Secondary value 2

Ʋ Channel 4

Temperature value

Transmission rate

31.25 Kbit/s

Current value

10 mA, ±0.5 mA

Additional output parameter - temperature
Processing is made via output signal HART multidrop, Profibus PA and Foundation Fieldbus
Range

-50 … +150 °C (-58 … +302 °F)

Resolution

1 °C (1.8 °F)

Accuracy

Ʋ in the range of 0 … +100 °C

(+32 … +212 °F)

±3 K

Ʋ in the range of -50 … 0 °C

(-58 … +32 °F) and +100 … +150 °C

(+212 … +302 °F)

typ. ±4 K

Input variable
Adjustment
Adjustment range of the min./max. adjustment relating to the nominal measuring range:

Ʋ Percentage value

-10 … 110 %

Ʋ Pressure value

-20 … 120 %

Adjustment range of the zero/span adjustment relating to the nominal measuring range:

Advertising