2 principle of operation – VEGA VEGABAR 66 Foundation Fieldbus User Manual

Page 9

Advertising
background image

9

3 Product description

VEGABAR 66 • Foundation Fieldbus

36740-EN-130321

2

1

13

10

5

4

3

6
7
8
9

11

12

14

Fig. 2: Layout of the type label (example)
1 Instrument type

2 Product code

3 Approvals

4 Electronics

5 Protection rating

6 Measuring range

7 Process and ambient temperature, process pressure

8 Material, wetted parts

9 Hardware and software version

10 Order number

11 Serial number of the instrument

12 ID numbers, instrument documentation

With the serial number, you can access the delivery data of the instru-

ment via www.vega.com, "VEGA Tools" and "serial number search".

In addition to the type label outside, you can also find the serial num-

ber on the inside of the instrument.

3.2 Principle of operation

The VEGABAR 66 is a suspension pressure transmitter for level

measurement in wells, basins and atmospherically open vessels.
For use in atmospherically closed vessels under vacuum, the instru-

ment is available with absolute pressure measuring ranges.

Sensor element is the CERTEC

®

measuring cell with robust, depend-

ent on the process fitting also front-flush, abrasion-resistant ceramic

diaphragm. The process pressure causes a capacitance change in

the measuring cell via the ceramic diaphragm. This change is con-

verted into an appropriate output signal and outputted as measured

value.
The CERTEC

®

measuring cell is also equipped with a temperature

sensor. The temperature value can be displayed via the display and

adjustment module as well as processed via the signal output (with

digital versions).

As a standard feature, the CERTEC

®

measuring cell is equipped with

a lateral, recessed seal.
Instruments with double seal have an additional front seal.

Application area

Functional principle

Seal concept