11 Supplement

VEGADIF 65 • Profibus PA

36129-EN-130417

Characterstic curve

Linear

Calibration position of the measuring cell Vertical, i.e. upright process component
Influence of the installation position to the

zero point

≤ 4 mbar

12)13)

A position-dependent zero-point shift can be corrected (see also chapter "Adjust parameter").
Position of the span in the measuring

range

Based on the zero point

Diaphragm material

316L, Alloy C276, gold rhodium plated, Monel

Filling oil

Silicone oil

Material, lateral flanges

316L

A position-dependent zero-point shift can be corrected (see also chapter "Adjust parameter").
Deviation determined according to the limit point method according to IEC 60770

14)

Applies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well as to analogue

current output 4 … 20 mA. Specifications refer to the set span. Turn down (TD) is the ratio nominal

measuring range/set span.
Deviation - All versions
The following applies to square root extracted characteristics: The accuracy data of VEGADIF 65

are entered with factor 0.5 in the accuracy calculation of the flow.
Deviation - Basic version
10 mbar, 30 mbar measuring cell

Ʋ Turn down 1 : 1

±0.15 % of the set span

Ʋ Turn down > 1 : 1

±0.15 % of the set span x TD

100 mbar measuring cell

Ʋ Turn down 1 : 1 to 4 : 1

±0.075 % of the set span

Ʋ Turn down > 4 : 1

±(0.012 x TD + 0.027) % of the set span

Measuring cells ≥ 500 mbar

Ʋ Turn down 1 : 1 to 15 : 1

±0.075 % of the set span

Ʋ Turn down > 15 : 1

±(0.0015 x TD + 0.053) % of the set span

Deviation - Chemical seal versions
100 mbar measuring cell

Ʋ Turn down 1 : 1 to 4 : 1

±0.075 % of the set span + influence of the chemical

seal

Ʋ Turn down > 4 : 1

±(0.012 x TD + 0.027) % of the set span + influence of

the chemical seal

Measuring cells ≥ 500 mbar

Ʋ Turn down 1 : 1 to 15 : 1

±0.075 % of the set span + influence of the chemical

seal

Ʋ Turn down > 15 : 1

±(0.0015 % x TD + 0.053 %) of the set span + influence

of the chemical seal

12)

Max. value with horizontal process component. Specification applies to the basic version without chemical seal.

The value doubles with instruments with inert oil.

13)

14)

Incl. non-linearity, hysteresis and non-repeatability.