VEGA VEGABAR 51 4 … 20 mA_HART User Manual

Page 53

Advertising
background image

53

10 Supplement

VEGABAR 51 • 4 … 20 mA/HART

36713-EN-130321

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psi)

Determination of characteristics

Limit point adjustment according to IEC 61298-2

Characterstic curve

Linear

Reference installation position

upright, diaphragm points downward

Influence of the installation position

depending on the chemical seal version

Deviation determined according to the limit point method according to IEC 60770

11)

Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to ana-

logue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal

measuring range/set span.
Deviation

Ʋ Turn down 1 : 1

< 0.2 %

Ʋ Turn down up to 5 : 1

< 0.2 %

Ʋ Turn down up to 10 : 1

< 0.3 %

Influence of the product or ambient temperature
Thermal change zero signal and output span
Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to ana-

logue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal

measuring range/set span.
Thermal change zero signal and output span, reference temperature 20 °C (68 °F):

Ʋ In the compensated temperature

range 0 … +100 °C (+32 … +212 °F)

< 0.05 %/10 K x TD

Ʋ Outside the compensated tempera-

ture range

typ. < 0.05 %/10 K x TD

Thermal change, current output
Applies also to the analogue 4 … 20 mA current output and refers to the set span.
Thermal change, current output

< 0.05 %/10 K, max. < 0.15 %, each with -40 … +80 °C

(-40 … +176 °F)

0,15 %

-0,15 %

-40°C

-20°C

20°C

40°C

60°C

80°C

Fig. 28: Thermal change, current output

11)

Incl. non-linearity, hysteresis and non-repeatability.

Advertising