VEGA VEGABAR 51 4 … 20 mA_HART User Manual
Page 57
57
10 Supplement
VEGABAR 51 • 4 … 20 mA/HART
36713-EN-130321
Ʋ Diameter approx.
8 mm (0.315 in)
Ʋ Colour
Blue
Cable entry/plug
17)
Ʋ External housing
– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5
or:
– 1␣x plug (depending on the version), 1␣x blind stopper
M20␣x␣1.5
Spring-loaded terminals for wire cross-
section up to
2.5 mm² (AWG 14)
Display and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Adjustment elements
4 keys
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the sensor without cover IP 40
Material
Ʋ Housing
ABS
Ʋ Inspection window
Polyester foil
Voltage supply
Operating voltage
Ʋ Non-Ex instrument
14 … 36 V DC
Ʋ Ex-ia instrument
14 … 30 V DC
Ʋ Exd instrument
18 … 36 V DC
Operating voltage with illuminated display and adjustment module
Ʋ Non-Ex instrument
20 … 36 V DC
Ʋ Ex-ia instrument
20 … 30 V DC
Ʋ Exd instrument
20 … 36 V DC
Permissible residual ripple
Ʋ < 100 Hz
U
ss
< 1 V
Ʋ 100 Hz … 10 kHz
U
ss
< 10 mV
Load
see diagram
17)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.