6 Setup

VEGASWING 66 • - two-wire

43758-EN-130322

Empty signal

Full signal

1

2

3

16

I/mA

8

3,6

1,5

1,5

1,5

t/s

Fig. 37: Flow chart of the function test
1 Full signal

2 Empty signal

3 Fault message

Check if all three switching conditions occur in the correct sequence

and the stated time period. If this is not the case, there is a fault in the

measuring system (see also the operating instructions manual of the

signal conditioning instrument). Keep in mind that connected instru-

ments are activated during the function test. By doing this, you can

check the correct function of the measuring system.
Note:

Keep in mind that the starting time t

A

of the voltage supply can extend

the time up to the first switching (e.g. VEGATOR 636: +1 s)

After releasing the button or after a brief line break.
The specified times apply with a tolerance of ±20 %.

Sensor

current -

Sensor

Level relay

A - overfill

protection

Signal lamp

A - Overfill

protection

Level relay

B - dry run

protection

Signal lamp

B - Dry run

protection

Fail safe

relay

Control

lamp

1. Fault sig-

nal
approx. 1.5 s
+ t

A

2)

< 3.6 mA

currentless

currentless

currentless

2. Empty

signal
(approx.

1.5 s)

approx.

8 mA

energized

currentless

energized

3. Full signal
(approx.

1.5 s)

approx.

16 mA

currentless

energized

energized

Test procedure

2)

Starting time of the voltage supply