VEGA VEGABAR 81 HART and accumulator pack - Operating Instructions User Manual

Page 33

Advertising
background image

33

6 Set up with the display and adjustment module

VEGABAR 81 • HART and accumulator pack

45026-EN-131107

Select the requested simulation size and adjust the requested value.
Caution:

During simulation, the simulated value is outputted as 4 … 20 mA cur-

rent value and digital HART signal.
Push the [ESC] key to deactivate the simulation.
Information:

The simulation is terminated automatically 60 minutes after the last

key has been pushed.

By entering a 4-digit PIN, you protect the sensor data against unau-

thorized access and unintentional modification.
In this menu item, the PIN is displayed or edited as well as modified.

However, it is only available when the adjustment is released in the

menu "Setup/ Lock/release adjustment ".

In delivery status, the PIN is "0000".

In this menu item, you adjust the internal clock of the sensor. There is

no adjustment to summer/winter time.

With a reset, certain parameter adjustments carried out by the user

are reset.

The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of

shipment from the factory incl. the order-specific settings. A user-pro-

grammable linearization curve as well as the measured value memory

will be deleted.

Additional settings - PIN

Additional adjustments -

Date Time

Additional adjustments

- Reset

Advertising