11 Supplement

VEGABAR 81 • HART and accumulator pack

45026-EN-131107

The specifications are only an overview and refer to the measuring cell. Limitations due to the

material and version of the process fitting as well as the selected pressure type are possible. The

specifications on the nameplate apply.

Nominal range

Overload capacity, max.

pressure

Overload capacity, min. pres-

sure

Gauge pressure
0 … +5 psig

+25 psig

-14.5 psig

0 … +15 psig

+75 psig

-14.5 psig

0 … +30 psig

+120 psig

-14.5 psig

0 … +150 psig

+450 psig

-14.5 psig

0 … +300 psig

+600 psig

-14.5 psig

0 … +600 psig

+1200 psig

-14.5 psig

0 … +1500 psig

+3000 psig

-14.5 psig

0 … +3000 psig

+6000 psig

-14.5 psig

0 … +9000 psig

+18000 psig

-14.5 psig

0 … +15000 psig

+22500 psig

-14.5 psig

-14.5 … 0 psig

+75 psig

-14.5 psig

-14.5 … +20 psig

+120 psig

-14.5 psig

-14.5 … +150 psig

+1200 psig

-14.5 psig

-14.5 … +300 psig

+1200 psig

-14.5 psig

-14.5 … +600 psig

+1200 psig

-14.5 psig

-3 … +3 psig

+25 psig

-14.5 psig

-7 … +7 psig

+75 psig

-14.5 psig

Absolute pressure
0 … +15 psi

+75 psi

0 psi

0 … +30 psi

+120 psi

0 psi

0 … +150 psi

+1200 psi

0 psi

0 … +300 psi

+1200 psi

0 psi

Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1

Ʋ Temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psi)

Determination of characteristics

Limit point adjustment according to IEC 61298-2

Characterstic curve

Linear

Reference installation position

upright, diaphragm points downward

Influence of the installation position

depending on the chemical seal version