VEGA VEGABAR 81 HART and accumulator pack - Operating Instructions User Manual

Page 53

Advertising
background image

53

11 Supplement

VEGABAR 81 • HART and accumulator pack

45026-EN-131107

Deviation (according to IEC 60770)
Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to the

analogue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal

measuring range/set span.

Accuracy class

Non-linearity, hysteresis and re-

peatability with TD 1 : 1 up to 5 : 1

Non-linearity, hysteresis and re-

peatability with 5 : 1

0.2 %

< 0.2 %

< 0.04 % x TD

Influence of the product or ambient temperature
Thermal change zero signal and output span
Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to the

analogue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal

measuring range/set span.

Average temperature coefficient,

zero signal

In the compensated tem-

perature range 10 … +70 °C

(+50 … +158 °F)

Outside the compensated tem-

perature range

Turn down 1 : 1

< 0.05 %/10 K

typ. < 0.05 %/10 K

Turn down 1 : 1 up to 5 : 1

< 0.1 %/10 K

-

Turn down up to 10 : 1

< 0.15 %/10 K

-

Thermal change, current output
Applies also to the analogue 4 … 20 mA current output and refers to the set span.
Thermal change, current output

< 0.05 %/10 K, max. < 0.15 %, each with -40 … +80 °C

(-40 … +176 °F)

0,15 %

-0,15 %

-40°C

-20°C

20°C

40°C

60°C

80°C

Fig. 21: Thermal change, current output

Additional temperature influence through chemical seal
The specifications refer to diaphragm material 316L as well as isolating liquid silicone oil. They are

only used for estimation. The actual values depend on the diameter, material and strength of the

diaphragm as well as the isolating liquid. They are available on request.
Temperature coefficient of the chemical seal in mbar/10 K with

Ʋ Flange DN 50 PN 40, Form C,

DIN 2501

1.2 mbar/10 K

Advertising