Quick start

VEGAPULS 62 standpipe version • Modbus and Levelmaster protocol

41363-EN-121011

Quick start

The quick start enables a quick setup with many applications. You

can find further information in the respective chapters of the operating

instructions manual.
1. In case of turbulences or strong product movement in the vessel,

long standpipe antennas should be fastened to the vessel wall.

2. The measurement is only possible inside the tube, the standpipe

antenna must hence reach up to the requested min. level

For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specifica-

tions on the type label.

2. Connect the instrument according to the following illustration

+

+

power supply

MODBUS

D0

D1

IS GND

USB

1

2

1

3 4 5

2

off on

( )

(

)

(-)

(-)

Fig. 1: Connection compartment
1 USB connection

2 Slide switch for integrated termination resistor (120 Ω)

For further information see chapter "Connecting to power supply".

The instrument is preset by default with the suitable parameters for

"Application", "Min. adjustment" and "Max. adjustment".
For adaptation of the parameter "Medium" you have to proceed as

follows:
1. Go via the indicating and adjustment module to the menu "Setup".

2. In the menu item "Medium"you select the medium of your applica-

tion, for example "Solvent".

1. In the menu "Additional settings", menu item "Damping" you have

to adjust the requested damping of the output signal.

2. Select the output characteristics in the menu item "Current out-

put".

The quick start is then finished. For further information see chapter

"Parameter adjustment".

Mounting

Connect electrically

Set parameters

Further steps