VEGA VEGAFLEX 86 Profibus PA User Manual

Page 89

Advertising
background image

89

11 Supplement

VEGAFLEX 86 • Profibus PA

44232-EN-130910

Gas phase

Temperature

Pressure

10 bar

(145 psig)

50 bar

(725 psig)

100 bar

(1450 psig)

200 bar

(2900 psig)

400 bar

(5800 psig)

Steam (satu-

rated steam)

100 °C/212 °F -

-

-

-

-

180 °C/356 °F 2.1 %

-

-

-

-

264 °C/507 °F 1.44 %

9.2 %

-

-

-

366 °C/691 °F 1.01 %

5.7 %

13.2 %

76.0 %

-

Characteristics and performance data
Measuring cycle time

< 500 ms

Step response time

4)

≤ 3 s

Max. filling/emptying speed

1 m/min

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
For the process conditions, please also note the specifications on the type label. The lowest value

always applies.
The measurement error from the process conditions is in the specified pressure and temperature

range of below 1 %.
Vessel pressure relating to the flange

nominal stage

see supplementary instructions manual "Flanges ac-

cording to DIN-EN-ASME-JIS"

Process pressure

-1 … +400 bar/-100 … +40000 kPa

(-14.5 … +5800 psig), depending on the process fitting

Process temperature

-196 … +280 °C (-321 … +536 °F)

1

2

0°C

(32°F)

-40°C

(-104°F)

65°C

(149°F)

80°C

(176°F)

280°C

(536°F)

250°C

(482°F)

190°C

(374°F)

3

4

-196°C

(-321°F)

-80°C

(-112°F)

55°C

(131°F)

Fig. 63: Ambient temperature - process temperature (version -196 … +280 °C/-321 … +536 °F)
1 Ambient temperature

2 Process temperature (depending on the seal material)

3 Maximum permissible temperature - standard

4 Limited temperature range - plastic housing and stainless steel housing, electropolished

4)

Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk

solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2).

Advertising