Quick start – VEGA VEGAFLEX 86 4 … 20 mA_HART two-wire Coax probe User Manual

Page 2

Advertising
background image

2

Quick start

VEGAFLEX 86 • 4 … 20 mA/HART two-wire

42284-EN-130612

Quick start

The quick start procedure enables a quick setup with many applica-

tions. You can find further information in the respective chapters of the

operating instructions manual.

Distance from the metallic vessel wall > 300 mm. Distance from

non-metallic vessel wall > 500 mm. The probe must not touch any

installations or the vessel wall.

For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specifica-

tions on the type label.

2. Connect the instrument according to the following illustration

5

1

2

+

( )

(-)

6 7 8

4...20mA

2

3

4

1

Fig. 1: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output

2 For display and adjustment module or interface adapter

3 For external display and adjustment unit

4 Ground terminal for connection of the cable screen

See also chapter "Connecting to power supply"

For standard applications we recommend selecting the "Quick setup"

in the display and adjustment module.

1. In this menu item you can select the application. You can choose

between level and interface measurement.

2. In the menu item "Medium - Dielectric constant" you can define

the medium type (medium).

3. Carry out the adjustment in the menu items "Min. adjustment" and

"Max. adjustment".

Mounting

Electrical connection

Set parameters

Advertising