VEGA VEGAFLEX 86 4 … 20 mA_HART two-wire Coax probe User Manual

Page 45

Advertising
background image

45

6 Set up with the display and adjustment module

VEGAFLEX 86 • 4 … 20 mA/HART two-wire

42284-EN-130612

The default setting for the displayed value 2 is the electronics tem-

perature.

The optionally integrated background lighting can be adjusted via the

adjustment menu. The function depends on the height of the supply

voltage, see "Technical data".

The lighting is switched off in the delivery status.

In this menu item, the device status is displayed.

The respective min. and max. measured value is saved in the sen-

sor. The two values are displayed in the menu item "Peak values,

distance".
If you have selected interface measurement under the menu item

"Setup - Application", the peak values of the interface measurement

are displayed in addition to the peak values of the level measurement.

In another window you can carry out a reset of the two peak values

separately.

The respective min. and max. measured values are saved in the

sensor. The two values are displayed in the menu item "Peak values,

measurement certainty".
The measurement can be influenced by the process conditions. In

this menu item, the measurement certainty of the level measurement

is displayed as percentage value. The higher the value, the more reli-

able the measurement. Values > 90 % indicate reliable measurement.
If you have selected interface measurement under the menu item

"Setup - Application", the peak values of the interface measurement

are displayed in addition to the peak values of the level measurement.

Display - Backlight

Diagnostics - Device

status

Diagnostics - Peak values

Distance

Diagnostics - Peak values

Measurement certainty

Advertising