VEGA VEGAFLEX 83 4 … 20 mA_HART two-wire Rod and cable probe - PFA insulated With SIL qualification User Manual

Page 57

Advertising
background image

57

7 Set up with the display and adjustment module

VEGAFLEX 83 • 4 … 20 mA/HART two-wire

44228-EN-130823

In this menu item you gain access to the protected area where

you can enter special parameters. In exceptional cases, individual

parameters can be modified in order to adapt the sensor to special

requirements.
Change the settings of the special parameters only after having con-

tacted our service staff.

The sensor offers the HART modes "Analogue current output" and

"Fix current (4 mA)". In this menu item you determine the HART mode

and enter the address with Multidrop mode.
In the mode "Fixed current output" up to 63 sensors can be operated

on one two-wire cable (Multidrop operation). An address between 0

and 63 must be assigned to each sensor.
If you select the function "Analogue current output" and also enter

an address number, you can output a 4 … 20 mA signal in Multidrop

mode.
With the mode "Fixed current (4 mA)" a fixed 4 mA signal is outputted

independently of the actual level.

The default setting is "Analogue current output" and the address 00.

The sensor is fix adjusted to the HART mode "Analogue current

output". This parameter cannot be modified.

The default setting is "Analogue current output" and the address 00.

In this menu, you read out the instrument name and the instrument

serial number:

Additional adjustments -

Special parameters

Additional adjustments -

HART mode

Additional adjustments -

HART mode

Info - Instrument name