VEGA VEGABAR 82 4 … 20 mA_HART - Operating Instructions User Manual

Page 75

Advertising
background image

75

11 Supplement

VEGABAR 82 • 4 … 20 mA/HART

45028-EN-131010

Ʋ Size of digits

W x H = 7 x 13 mm

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the housing without

cover

IP 40

Materials

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Interface to the external display and adjustment unit
Data transmission

digital (I²C-Bus)

Configuration, connection cable

4-wire, screened

Cable length max.

Ʋ Sensors with signal output

4 … 20 mA, 4 … 20 mA/HART

50 m

Ʋ Sensors with signal output Profibus

PA, Foundation Fieldbus

25 m

Interface to the slave sensor
Data transmission

digital (I²C-Bus)

Configuration, connection cable

4-wire, screened

Cable length max.

25 m

Integrated clock
Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Rate deviation max.

10.5 min/year

Measurement electronics temerature
Resolution

1 °C (1.8 °F)

Accuracy

±1 °C (1.8 °F)

Voltage supply
Operating voltage U

B

Ʋ Non-Ex instrument

9.6 … 35 V DC

Ʋ Ex-d instrument

9.6 … 35 V DC

Ʋ Ex-ia instrument

9.6 … 30 V DC

Ʋ Ex-d-ia instrument

15 … 35 V DC

Ʋ Ex-d-ia instrument with ship approval 15 … 35 V DC

Interpolation protection

Integrated

Permissible residual ripple - Non-Ex, Ex-ia instrument

Advertising