6 Set up with the display and adjustment module

VEGABAR 82 • Profibus PA

45031-EN-131107

To damp process-dependent measured value fluctuations, you can

set a time of 0 … 999 s in this menu item.
The damping applies to the level and interface measurement.

The default setting is a damping of 0 s.

In the menu item "Lock/unlock adjustment", you can protect the

sensor parameters against unauthorized modification. The PIN is

activated/deactivated permanently.
The following adjustment functions are possible without entering the

PIN:

Select menu items and show data

Read data from the sensor into the display and adjustment mod-

ule.

Caution:

With active PIN, adjustment via PACTware/DTM as well as other

systems is also blocked.
You can change the PIN number under "Additional adjustments -

PIN".

This menu item enables the setting of the requested national lan-

guage.

The following languages are available:

German

English

French

Spanish

Russian

Italian

Dutch

Portuguese

Polish

Setup - AI FB1 - Damping

Lock/release setup - Ad-

justment

Display - Language