GE Industrial Solutions MicroVersaTrip PM User Manual

Page 26

Advertising
background image

MicroVersaTrip Plus™ and MicroVersaTrip PM™ Trip Units

Chapter 2. Setup Mode

16

Long-Time Pickup

The first setup-mode display is always the long-time
pickup setpoint, as illustrated in Figure 10. This set
point establishes the breaker’s nominal ampere rat-
ing,

xLT, as a fraction of the rating plug value, xIn

(

xLT = LT multiplier x xIn). Press the

VALUE

key to

scroll through the available choices. Press

ENTER

to

store the desired set point.

Figure 10. Trip Unit display for long-time pickup.

The choices for UL breakers are 0.50 to 1.00 times
xln in steps of 0.05. The pickup value tolerance band
is 0% to +20% of the set point.

The choices for ANSI breakers are 0.50 to 1.10 times
xln, in steps of 0.05. The pickup value is defined for

−1

0% to +10% of the set point.

Figure 11 illustrates the long-time pickup settings.

Figure 11. Time-current curve illustrating long-time pickup.

Long-Time Delay

The Trip Unit display for long-time delay is illus-
trated in Figure 12. This function allows normal
momentary overloads without nuisance tripping.
The time delays at the lower limit of the bands at
600% of the long-time current setting,

xLT, are listed

in Table 11. Figure 13 illustrates the effect of this
delay on trip time. Press the

V A L U E

key to cycle

through the four choices of time-delay bands. Press

ENTER

to store the desired value.

Figure 12. Trip Unit display for long-time delay.

B

B

B

Baa

a

an

n

n

nd

d

d

d

D

D

D

Deee

ellllaa

a

ayyyy,,,, sssseee

eccc

c

1

2.4

2

4.9

3

9.8

4

20

Table 11. Lower-limit delays for long-time delay bands.

Figure 13. Time-current curve illustrating long-time delay.

Short-Time Pickup

The short-time pickup function establishes the cur-
rent at which short-time trip is activated. Short-time
pickup is coupled with long-time pickup and the
choices of pickup settings are from 1.5 to 9.0 times
the long-time setting,

xLT, in steps of 0.5 xLT. The

Trip Unit display is illustrated in Figure 14.

Advertising