3 conductivity measurement, 4 d.o. measurement, Conductivity measurement – Xylem ProLab 2000 User Manual

Page 102: D.o. measurement

Advertising
background image

What to do if...

ProLab 2000

102

ba75577e05

11/2013

Warning [TempErr]

Warning [ISEErr]

6.3

Conductivity measurement

Conductivity measuring

cell was not recognized

Error message, Error

(invalid

calibration)

CalClock flashes

6.4

D.O. measurement

D.O. sensor was not

recognized

Cause

Remedy

– Temperature difference

between measurement and
calibration greater than 2 K.

– Adjust the temperature of

the test sample

Cause

Remedy

– Electrode voltage outside

calibrated range

– Recalibrate

Cause

Remedy

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable control standard

– Use control standard,

0.01 mol/l KCl

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Sensor not connected

– Connect probe

– Cable broken

– Replace probe