4 settings for conductivity measuring cells, Settings for conductivity measuring cells – Xylem ProLab 2000 User Manual

Page 66

Advertising
background image

Operation

ProLab 2000

66

ba75577e05

11/2013

4.7.4

Settings for conductivity measuring cells

Overview

The following settings are possible for conductivity measuring cells:

 Measured parameter
 Reference temperature
 Temperature compensation
 TDS factor
 Calibration interval
 Measuring cell / cell constant

Settings

The settings are made in the measuring menu of the conductivity
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the <MENU/OK> key.
After completing the settings, switch to the measured value display with
<MODE>.

Menu item

Possible
setting

Description

Calibration /
Calibration interval

1 ... 999 d

Calibration interval for the
measuring cell (in days).
The meter reminds you to
calibrate regularly by the
flashing CalClock in the
measuring window.

Calibration /
Calibration record

-

Displays the calibration
record of the last calibration.

Calibration /
Calibration history

-

Displays the calibration
history of the last calibrations.

Reset

-

Resets all sensor settings to
the delivery condition (see
page 93).

Temp. comp. (TC) /
Reference temp.

20°C
25 °C

Reference temperature

This setting is only available
when the Conductivity or
Resistivity display is set.

Temp. comp. (TC) /
Compensation

nLF
lin
Off

Procedure for temperature
compensation (see page 65).

This setting is only available
when the Conductivity or
Resistivity display is set.