4 d.o. calibration, D.o. calibration – Xylem ProLab 2000 User Manual

Page 75

Advertising
background image

ProLab 2000

Operation

75

ba75577e05

11/2013

4.8.4

D.O. calibration

Why calibrate?

D.O. sensors age. This changes the slope of the D.O. sensor.
Calibration determines the current slope of the sensor and stores this
value in the instrument.

When to calibrate?

 After connecting another non-ID D.O. sensor
 When the CalClock flashes (after the calibration interval has

expired).

Calibration datasets

The ProLab 2000 administrates two sets of calibration data:

Sensors of different types can be calibrated separately from each other.
When a sensor of one type is calibrated, the calibration data of the
other types remain stored. The ProLab 2000 recognizes the type of the
connected sensor and automatically uses the correct calibration data.

Calibration procedures

The ProLab 2000 provides 2 calibration procedures:

 Calibration in water vapor-saturated air.

Use an air calibration vessel for calibration.

 Calibration via a comparison measurement (e.g. Winkler titration

according to DIN EN 25813 or ISO 5813). At the same time, the
relative slope is adapted to the comparison measurement by a
correction factor. When the correction factor is active, the [Factor]
indicator appears in the measuring window.

Stability control

During calibration, the stability control is automatically activated.

Display of calibration

data and download to

interface

You can have the data of the last calibration displayed (see page 87).
Subsequently, you can transmit the displayed calibration data to the
interface, e. g. to a printer or PC, with the <PRINT> key.

Note
The calibration record is automatically transmitted to the interface after
calibrating.

 Set 1 for the type, "Ox":

– 9009/6. or

– Ox1400Stir

 Set 2 for the type,

"FOx1400":

– LFOx1400

Advertising