2 measuring ranges, resolution, accuracy, 1 ph/orp, Measuring ranges, resolution, accuracy 7.2.1 – Xylem ProLab 2000 User Manual

Page 106: Ph/orp

Advertising
background image

Technical data

ProLab 2000

106

ba75577e05

11/2013

Stand March 3, 2010

Translation of the legally binding German version

7.2

Measuring ranges, resolution, accuracy

7.2.1

pH/ORP

Note
The measuring ranges and accuracy values specified here apply
exclusively to the meter. The accuracy of the sensors has also to be
taken into account.

Instrument safety

EC guideline 2006/95/EC
EN 61010-1
ANSI/UL 61010-1
CAN/CSA-C22.2 No. 61010-1

Radio data
transmission

EC guideline 1999/5/EC
EN 300 330-2
EN 50364
EN 60950-1

Climatic class

VDI/VDE 3540

IP protection class

EN 60529

Measuring ranges,

resolution

Variable

Measuring range

Resolution

pH

- 2.000 ... + 20.000
- 2.00 ... + 20.00

0.001
0.01

U [mV]

- 1999.9 ... + 1999.9
- 1999 ... + 1999

0.1
1

T [°C]

- 10.0 ... + 120.0

0.1

Manual

temperature input

Variable

Range

Increment

T

manual

[°C]

- 10 ... + 130

1

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

pH

± 0.003
± 0.01

+ 15 °C ... + 35 °C

U [mV] / range

- 999.9 ... + 999.9

± 0.2

+ 15 °C ... + 35 °C

- 1999 ... + 1999

± 1

+ 15 °C ... + 35 °C

T [°C] / temperature sensor

± 0.1

0 °C ... + 55 °C

Advertising