5 general errors, General errors – Xylem ProLab 2000 User Manual
Page 104
Advertising
What to do if...
ProLab 2000
104
ba75577e05
11/2013
CalClock flashes
6.5
General errors
Meter does not react
to keystroke
Time is lost
– The carbon dioxide pressure is
too high (> 1 bar)
– Measuring not possible
Cause
Remedy
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <On/Off> and
<SC> key simultaneously.
Cause
Remedy
– Buffer batteries empty
– Exchange buffer batteries