background image

pHotoFlex / pHotoFlex Turb

Operation

75

ba75491d04

07/2006

4.9.4

Transmitting data

The following table shows which data are transmitted to the interface in
which way:

Note
With the <PRT> key you output the data that is being shown on the dis-
play to the interface (displayed measured values, stored measurement
datsets, calibration record).

12

Unit of 11

! °C / °F
! nm

13

Measured value
status of 11

VALID, INVALID, UFL, OFL

Data

Operation / description

Current
measured value

! Press <PRT>.
! Simultaneously with every manual storage

process.

Stored measured
values

! Display stored dataset and press <PRT>.
! All datasets according to the filter criteria via

the RS232 download function
(see section 4.8.2).

Calibration record
(pH electrode)

! Calibration record of a pH electrode with

<PRT> (after calling up from the storage or at
the end of a calibration).

! Calibration record of a pH electrode at the

end of a calibration procedure is printed au-
tomatically.

Data

Description

Advertising