6 what to do if, 1 general errors, What to do if – Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 87: General errors, Chapter 6 w, 6what to do if
pHotoFlex / pHotoFlex Turb
What to do if...
87
ba75491d04
07/2006
6
What to do if...
6.1
General errors
Display, LoBat
Instrument does not
react to keystroke
RS232 interface
does not react
Error message,
Error
0, 8, 16, 16384
Cause
Remedy
– The batteries or accumulator
pack are largely depleted
– Insert new batteries
– Charge the accumulator pack
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <START/ENTER>
and <PRT> key
simultaneously.
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <START/ENTER>
and <PRT> key
simultaneously.
Cause
Remedy
– Instrument error
– Repeat measurement
– Meter defective,
send meter to WTW for repair
and quote the error number