4.13.1 Administrating user-defined programs with a terminal

program

Generally, a terminal program serves to establish a connection to a de-
vice on a data interface and to communicate with the device via a con-
sole on the display.
Thus, it is also possible to send command lines.
Terminal programs are available for different operating systems by dif-
ferent manufacturers. Windows (version 95 to XP) contains the "Hyper-
Terminal" terminal program. It is in the program menu under
accessories.
For more detailed information please refer to the user information of the
terminal program.

Note
The CD-ROM contains a configuration file for the HyperTerminal pro-
gram. By double-clicking the configuration file (*.ht), the HyperTerminal
with the required pre-settings for data exchange with the meter is
opened.
The pHotoFlex (Turb) can administrate the database of user-defined
programs via command lines.

Storing user-defined

programs

Enter a command line according to the following scheme in the terminal
program:
U.500#11,Program No,Model,Cuvette [mm],Wave length [nm],MRB,
MRE,E0,Slope,Resolution,Formula,Unit

Example:
U.500#11,900,Test,16,436,0.0,2.0,0.0,1.0,0.01,test,

mg/l

The individual data sections of the command line are separated by
commas. The dot "." has to be used as a decimal separator within a
data section.

1

With the aid of the AK 540/B interface cable, connect the
pHotoFlex (Turb) 540 to the serial interface (COM port) of the
PC (see section 8.1.1).

2

Make sure the pHotoFlex (Turb) is switched on.

3

Start the terminal program on the PC.

4

Configure the connection settings for the COM interface as
necessary.

5

Enter the command line.