E8361c, Uncorrected system performance 1 – Atec Agilent-E8364C User Manual

Page 35

Advertising
background image

35

E8361C

Uncorrected system performance 1

continued

Description Specification

Supplemental

information

Reflection tracking

Typical:

10 to 45 MHz

±1.5 dB

45 MHz to 20 GHz

±1.5 dB

20 to 40 GHz

±2.0 dB

40 to 50 GHz

±2.0 dB

50 to 67 GHz

±3.0 dB

67 to 70 GHz

±4.5 dB

Transmission tracking

3

Typical:

10 to 45 MHz

±1.5 dB

45 MHz to 20 GHz

±1.5 dB

20 to 40 GHz

±2.0 dB

40 to 50 GHz

±2.0 dB

50 to 67 GHz

±3.0 dB

67 to 70 GHz

±4.5 dB

Crosstalk

4

- standard

10 to 45 MHz

2

63 dB

45 to 500 MHz

87 dB

500 MHz to 2 GHz

110 dB

2 to 10 GHz

105 dB

10 to 24 GHz

111 dB

24 to 30 GHz

106 dB

30 to 40 GHz

104 dB

40 to 45 GHz

98 dB

45 to 50 GHz

100 dB

50 to 60 GHz

97 dB

60 to 67 GHz

94 dB

67 to 70 GHz

2

94 dB

Crosstalk

4

- Option 014

Typical (for Option 080 enabled

5

)

10 to 45 MHz

2

63 dB

63 dB

45 to 500 MHz

87 dB

87 dB

500 MHZ to 2 GHz

110 dB

110 dB

2 to 10 GHz

105 dB

105 dB

10 to 24 GHz

111 dB

111 dB

24 to 30 GHz

104 dB

104 dB

30 to 40 GHz

102 dB

102 dB

40 to 45 GHz

96 dB

96 dB

45 to 50 GHz

98 dB

98 dB

50 to 60 GHz

95 dB

95 dB

60 to 67 GHz

90 dB

90 dB

67 to 70 GHz

2

90 dB

90 dB

Crosstalk - Option 014 and UNL

Typical (for Option UNL and

Option 014 with 080 enabled

5)

10 to 45 MHz

2

63 dB

63 dB

45 to 500 MHz

87 dB

87 dB

500 MHz to 2 GHz

110 dB

110 dB

2 to 10 GHz

104 dB

104 dB

10 to 24 GHz

108 dB

108 dB

24 to 30 GHz

101 dB

101 dB

30 to 40 GHz

99 dB

99 dB

40 to 45 GHz

92 dB

92 dB

45 to 50 GHz

94 dB

94 dB

50 to 60 GHz

91 dB

91 dB

60 to 67 GHz

84 dB

84 dB

67 to 70 GHz

2

84 dB

84 dB

1. Specifications apply over environment temperature of 23 °C ± 3 °C, with less than
1 °C deviation from the calibration temperature.
2. Typical performance.
3. Transmission tracking performance noted here is normalized to the insertion loss

characteristics of the cable used, so that the indicated performance is independent

of cable used.

4. Measurement conditions: Normalized to a thru, measured with two shorts, 10-
Hz

IF bandwidth, averaging factor of 16, alternate mode, source power set to

the lesser of the maximum power out or the maximum receiver power.
5. 0 Hz offset.

Advertising