Microwave pna series, General information – Atec Agilent-E8364C User Manual

Page 47

Advertising
background image

47

Microwave PNA Series

General information

Description Supplemental

information

System IF bandwidth range

1 Hz to 40 kHz, nominal

RF connectors
E8362C

3.5 mm (male), 50 , (nominal), center pin recession flush to .002 in. (characteristic)

E8363/4C

2.4 mm (male), 50 , (nominal), center pin recession flush to .002 in. (characteristic)

E8361C

1.85 mm (male), 50 , (nominal), center pin recession flush to .002 in. (characteristic)

Display

8.4 in diagonal color active matrix LCD; 1024 (horizontal) x 768 (vertical) resolution;

60 Hz refresh rate

A display is considered faulty if:

* A complete row or column of "stuck" or "dark" pixels.

* More than six "stuck on" pixels (but not more than three green) or

more than 0.002% of the total pixels are within the LCD specifications.

* More than twelve "dark" pixels (but no more than seven of the same color)

or more than 0.004% of the total pixels are within the LCD specifications.

* Two or more consecutive "stuck on" pixels or three or more consecutive

"dark" pixel (but no more than one set of two consecutive dark pixels)

"Stuck on" of "dark" pixels less than 6.5 mm apart (excluding consecutive pixels)

Display range
Magnitude

±200 dB (at 20 dB/div), max

Phase

±500°,

max

Polar

10 pico units, min; 1000 units, max

Display resolution
Magnitude

0.001

dB/div,

min

Phase

0.01°/div,

min

Marker resolution
Magnitude

0.001 dB, min

Phase

0.01°,

min

Polar

0.01 mUnit, min; 0.01°, min

CPU Intel

®

1.1 GHz Pentium

®

M with 1 GByte RAM

Line power

(single phase)

Frequency

50/60/400 Hz for 100 to 120 V, 50/60 Hz for 220 to 240 V (Power supply is auto switching.)

Max

350

Watts

General environmental
EMC

Complies with European EMC directive 2004/108/EC

IEC/EN

61326-1:2005

• CISPR Pub 11 Group 1, class A

• AS/NZS CISPR II:2004

ICES/NMB-001

Safety

Complies with European Low Voltage Directive 2006/95/EC

IEC/EN

61010-1:2001

• Canada: CSA C22.2 No. 61010-1:2004

• USA: UL 61010-1:2004

Operating environment
Temperature

0 to +40 °C; Instrument powers up, phase locks, and displays no error messages

within this temperature range. (Except for ‘source unleveled’ error message that

may

occur at temperature outside the specified performance temperature range of 25 °C, ± 5 °C.)

Error-corrected temperature range

System specifications valid from 23 °C, ± 3 °C, with less than 1 °C deviation from the

calibration

temperature

Relative humidity

Type-tested, 0 to 95% at 40 °C, non condensing

Altitude

0 to 4600 m (15,000 ft)