E8361c, Test port input, Dynamic accuracy (specification) – Atec Agilent-E8364C User Manual

Page 45: Continued

Advertising
background image

45

E8361C

Test port input

continued

Dynamic accuracy (specification)

1

Applies to input ports 1 and 2, accuracy of the test
port input power reading relative to the reference
input power level. Also applies to the following
conditions:

• IF bandwidth = 10 Hz

Note: If the power is set above maximum specified
leveled power, the test port output signal may show
non-linear effects that are dependent on the DUT.

Magnitude

0.01

0.1

1

10

-120

-110

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Testport power (dBm)

E8361C 014

-10 dBm (0.500-5 GHz)

-20 dBm (0.500-5 GHz)
-30 dBm (0.500-5 GHz)

-40 dBm (0.500-5 GHz)

Accuracy (dB)

Phase

0.1

1

10

100

-120

-110

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Testport power (dBm)

E8361C 014

-10 dBm (0.500-5 GHz)

-20 dBm (0.500-5 GHz)
-30 dBm (0.500-5 GHz)

-40 dBm (0.500-5 GHz)

Accuracy (degrees)

1. Dynamic accuracy is verified with the following measurements: compression over

frequency, IF linearity at a single frequency of 1.195 GHz and and a reference level

of -20 dBm for an input power range of 0 to -120 dBm.