Andersen Windows & Doors AN74I User Manual

Page 10

Advertising
background image

10

OWNER’S MANUAL

After the initial purge, the screen to the right will appear.
If the operator breaks the ampoule and closes the door
without selecting a cycle length, the sterilizer will beep
after 5 seconds to remind the operator to select the appro-
priate cycle length. The operator should break the An-
prolene ampoule, close the sterilizer door, remove the key,
and select the desired cycle length.



Once the sterilization cycle has been initiated, the display
will show the cycle length chosen and the time remaining
in the sterilization cycle.









At the end of the sterilization cycle, the unit will begin a 2
hour sterilization liner bag ventilation (purge) cycle to
remove the residual Ethylene Oxide. The display will
show the time remaining in the purge cycle. During this
time the purge pump and cabinet ventilation pumps will
run alternating two minute cycles.

At the end of the 2 hour liner bag ventilation (purge) cycle,
the machine will beep once and instruct the user to
UNLOAD STERILIZER, signifying the end of the cycle.
The sterilizer may now be safely unloaded.

Note - Additional Aeration Option:
The ventilation pump and purge pump continue running
alternating two minute cycles until the door is opened and
the EXIT button is pressed. If needed, items may be left
in the machine for further aeration after the cycle has
ended by simply leaving the items inside the sterilizer.
The count-up timer indicates the additional aeration time.

Once the cycle has been concluded by opening the door
and pressing the EXIT button, both pumps will stop run-
ning and the sterilizer will return to the standby screen.

12 HOUR CYCLE

STERILIZING

11:24:15 REMAINING

VENTILATING BAG

01:59:55 REMAINING

UNLOAD STERILIZER

00:08:54

EXIT

AN74 I

ANPROLENE STERILIZER

START

BREAK AMPOULE

CLOSE & LOCK DOOR

SELECT CYCLE LENGTH

24 HOUR 12 HOUR

12 HOUR CYCLE

STERILIZING

11:24:15 REMAINING

When 12 hour cycle length
selected

Continued from previous page

Diagrammatic Instructions for Normal Operation, continued

When EXIT pressed

UNLOAD STERILIZER

01:29:30

EXIT

© Andersen Products Ltd

Advertising
This manual is related to the following products: