Test Vent
pump

When left
button held for
3 seconds

Flowchart for Technical and Diagnostic Screens

AN74 I

ANPROLENE STERILIZER

START

DIAGNOSTICS

NO ABATOR

DIAGNOSTICS

ABATOR ATTACHED

PURG V. 0.0 (>5.00)
VENT V. 0.00 (>0.25)
BACK P. 0.0 (<15.0)
EXIT FAN 102% (>70)

PURG V. 25.3 (>5.00)
VENT V. 0.00 (>0.25)
BACK P. 6.4 (<15.0)
EXIT ABATOR ON

PURG V. 0.0 (>5.00)

VENT V. 1.51 (>0.25)
BACK P. 8.2 (<15.0)

EXIT ABATOR OFF

PURG V. 25.3 (>5.00)
VENT V. 0.00 (>0.25)
BACK P. 6.4 (<15.0)
EXIT FAN 102% (>70)

If Abator
attached

If no Abator
attached

PURG V. 0.0 (>5.00)
VENT V. 0.94 (>0.25)
BACK P. 8.2 (<15.0)
EXIT FAN 102% (>70)

PRESS ABATOR TO

INSTALL CARTRIDGE,

OTHERWISE PRESS DIAG

ABATOR DIAG

PURG V. 0.0 (>5.00)

VENT V. 0.00 (>0.25)
BACK P. 0.0 (<15.0)
EXIT FAN 102% (>70)

These screens are used by authorized service technicians and installers only, to check the vacuum created
by the abator pump, ventilation pump and purge pump. The screens are used for diagnostic and installation
purposes and would not normally be used by the operator of the machine.