Andersen Windows & Doors AN74I User Manual

Page 66

Advertising
background image

66

OWNER’S MANUAL

12 (or 24) HOUR CYCLE

STERILIZING

1:24:15 REMAINING

VENTILATING BAG

PURGE PUMP FAILURE

24:00:00 REMAINING

UNLOAD STERILIZER

00:00:44

EXIT

VENTILATING BAG

0:23:55 REMAINING

If a PURGE PUMP FAILURE
occurs during the ventilation
cycle, an additional 24 hours
will automatically be added to
the time remaining

The PURGE PUMP FAILURE message will
remain on the screen. At the end of the ex-
tended cycle, the operator should: (1) open the
door; (2) remove the Velcro strap and open the
mouth of the bag; and (3) close and lock the
door.

Continue ventilation for an additional 2 hours.
Open the door and press EXIT. Then remove
the items from the cabinet as you normally
would.

ITEMS IN THE LINER BAG MAY BE
CONSIDERED STERILE

PURGE PUMP FAILURE

REFER TO ERROR

SECTION OF MANUAL

Flowchart for Purge Pump Failure During Ventilation Cycle

Refer to the Error and Troubleshooting sections
of the manual for information about repairing the
sterilizer

To reset the machine, shut off the machine power.
Then, while holding both buttons, turn on the
power. Release buttons after the display shows
characters.

AN74 I

ANPROLENE STERILIZER

START

PURGE PUMP FAILURE

OPEN DOOR

REMOVE VELCRO STRAP

CLOSE DOOR

RESET

PURGE PUMP FAILURE

01:59:59 REMAINING

Count up timer