Flowchart for Pre-Cycle Error Messages

COOLING FAN FAILURE indicates the cooling fan in the top did

not reach the required operating speed (70% of rated speed)

PURGE SENSOR, VENT SENSOR or PRESSURE SENSOR

FAILURE indicates the offset of the specified sensor was out of range
(> =0.1 AND <= 0.4) prior to starting the ventilation pump.

PURGE PUMP FAILURE indicates the purge sensor or the purge

pump failed or the purge tubing became disconnected.

VENT PUMP FAILURE indicates the ventilation tubing is blocked

or disconnected, the ventilation sensor failed or the ventilation pump
failed.

ABATOR FAILURE indicates the exhaust tubing blocked, discon-

nected or the abator pump failed.

BACK PRESSURE HIGH indicates the exhaust tubing is blocked.
EXIT returns to the startup screen

PURGE SENSOR FAILURE

VENT SENSOR FAILURE

PRESS SENSOR FAILURE

PURGE PUMP FAILURE

VENT PUMP FAILURE

ABATOR FAILURE

REFER TO ERROR

SECTION OF MANUAL

EXIT

250 ABATOR CYCLES

REMAINING

SELF TEST

OK

12,000 PUMP HOURS

FAN 94 % RPM

Go to normal operating screens

COOLING FAN FAILURE

CHECK REAR FAN

OPERATION & FILTER

EXIT BYPASS

BACK PRESSURE HIGH

CHECK EXHAUST

TUBING & FITTINGS

EXIT BYPASS

Possible Error Messages: