Burkert Type 8055 User Manual
Page 33
SE 56
33
210_EN_BU_4_3_5X.d
oc
MENU 8.DISPLAY
(POS. 8.3) Display contrast set
[CONTRAST=X]
Display visual contrast set. The contrast can change according to the room temperature. The set
values are from 0 to 15. The entered value has effect only when leaving the function itself.
Contrast also can be set from visualization pages (pag. 17) pushing the key for 8
second or more. In this way the contrast set that will be visualized at release of the key.
(POS. 8.12-8.13) Conversion factor for flow rate totaliser
[EUR/dm³+ =X]
Set the value of conversion/currency for direct totalise( positive ). There are three set fields for this
parameter, from left to right: 1) monetary token, 2 ) default/personalized monetary token, 3 )
conversion coefficient. For the selection setting the cursor over the field to modify. The mode set of
monetary token could be two:
1.
Choice of one of the 7 predetermined monetary tokens (standard ISO 4217-REV81):
EUR = Eur ; USD = USA ; dollar CAD = Canadian dollar ; AUD = Australian dollar ; GPB =
English pound ; CHF = Swissfranc ; JPY = Japanese yen.
2.
Choice three free characters (number or letter ) . To change the characters , the cursor has to
be positioned on the symbol “/” (field N. 2)
MENU 9.DATA LOGGER
(POS. 9.1) Date and time set
[· = dd/mm/yy hh:mm]
Date and time set. If the real time clock optional module is present, then the time setting is kept also
when the power supply is off, otherwise it is frozen till the power supply is back. For example, if the
power supply has been off for one hour, when switched on the instrument will be one hour late. The
calendar is valid till year 2091. N.B.: date and time are visualized on display only if the data logger is
enable (Pos 9.2).
MENU 10. DIAGNOSTIC
(POS. 10.1) Meter “calibration”
[CALIBRATION]
Enable the calibration of the meter. The activation of this function happens pressing the key
during the visualization of the function. Will be visualized the following question: " EXECUTE?" press
for more of two second the key to proceed . Press any other key to delete the operation
(POS. 10.2) “Autotest” function enable
[SELF TEST]
Meter autotest function. This function stops the normal functions of the meter and performs a
complete test cycle on the measure input circuits and on the excitation generator. To activate this
function, after select it, push key , at the question: “EXECUTE?” push the key
For start autotest, or any other key for delete operation. The result of the test is shown on the
display. At the end of operation will have visualized one of visualization page. This function is
automatically performed when switching on the device.
(POS. 10.3) Flow rate simulation
[SIMULATION]
Flow rate simulation enabling. With this function it is possible to generate an internal signal that
simulates the flow rate, allowing the outputs and all the connected instruments test. After enabling it,
the flow rate simulation can be:
- set: by pushing the key from one of four visualization pages
- started: by pushing the key after set it
- finished: by pushing the key from visualization pages and then by pushing the key .
N.B.: the enable of flow rate simulation disable the contrast regulation with the key (Pos. 8.3)