GROUNDING INSTRUCTIONS

CONVERTER POWER SUPPLY

q

Before connecting the power supply, verify that
the mains voltage falls between the limits

indicated on the tag plate

q

ATTENTION: the converters on dc power

supply line are not protected against the
inversions of polarity.

q

For the wiring use only approved conductors,
with fireproof properties.

q

The power supply line must be equipped with
an external protection for current overload (fuse

or automatic line breaker with limiting capacity
not greater than 10 A).

q

In the proximity of the instrument Provide a
circuit breaker that must be easily accessible

from the operator and clearly identified.

NOTE: characteristics of meter’s power supply, see

page 4

Wall version

Panel version

For the correct operation of the meter it’s NECESSARY that sensor and liquid are

equipotential, so ALWAYS connect sensor and converter to the ground