Installation and wiring

Type 8045

Horizontal mounting

Correct

Incorrect

Vertical mounting

Correct

Incorrect

flow direction

Fig. 6: Filling of the pipe

- Prevent the formation of air bubbles in the pipe in the section

around the device (see Fig. 7).

- Always mount the device upstream a possible injection point

in the pipe of a high-conductivity fluid (for example: acid, base,
saline,...).

Respect the following additional mounting conditions to ensure
that the measuring device operates correctly:

- Preferably install the device at a 45° angle to the horizontal centre of

the pipe to avoid having deposits on the electrodes and false meas-
urements due to air bubbles (see Fig. 5);

45°

45°

Fig. 5: Mounting angle on the pipe

- Ensure that the pipe is always filled in the section around the

device (see Fig. 6).

- When mounting vertically ensure that the flow direction is in an

upward direction (see Fig. 6).

English