Around the device (see fig. 7) – Burkert Type 8045 User Manual

Page 14

Advertising
background image

14

Installation and wiring

Type 8045

Correct

flow direction

Incorrect

Correct

Incorrect

Fig. 7: Air bubbles within the pipe

If necessary, use a flow conditioner to improve measurement
precision.

7.3.

installation into the pipe of a
8045 with a G2'' nut

In order to ensure a high accuracy of the measurements and
good stability of the „flow zero“ point, install the device into
the processed medium at least 24 H before calibration.

Install the S020 fitting into the pipe taking into account the rec-
ommendations in chap. 7.2.

Check that there is a seal on the fitting and that it is not damaged.
Replace the seal if necessary.

Check that there is a seal (see mark 6, Fig. 8) on the flow sensor.

Insert the nut (see mark 3, Fig. 8) on the fitting.

Insert the snap ring (mark 2 Fig. 8) into the groove (mark 5 Fig.
8
).

Position the arrow on the side of the device in the direction of the
flow: the totalizers will increment.

Insert the device (mark 1 Fig. 8) into the fitting.

Tighten the nut (mark 3, Fig. 8) by hand on the device.

English

Advertising