Error Error text

#Comment

Outputs

Input

No.

Remedy text

status

status

1

I1 - Power on
OK

Power on has activated.

Active

Active

2

I2 - Add-on Module
Applied

A new module has been added to the system.

Active

Active

3

I3 - Add-on Module

An add-on module is bad or has been removed.

Install

This can also be an internal add-on module.

Active

Active

4

I4 - Param. corrected

A less vital parameter in the converter has been

OK

replaced by its default value.

Active

Active

20

W20 - Totalizer 1

During initialization, the check of the saved totalize

Reset manually

value failed. The saved totalizer value is not reliable.
Reset the totalizer value manually to rely on readings.

Active

Active

20

W20 - Totalizer 2

During initialization, the check of the saved totalize

Reset manually

value failed. The saved totalizer value is not reliable.
Reset the totalizer value manually to rely on readings.

Active

Active

21

W21 - Pulse overflow

Actual flow is too big compared with pulse width and

Reduced

Adjust pulse settings

mass/pulse.

pulse width

Active

22

W22 - Batch timeout

Duration of batching has exceeded a predefined

Batch out-

Check installation

maximum time.

put on zero

Active

23

W23 - Batch overrun

Batch quantity has exceeded a predefined maximum

Batch out-

Check installation

overrun mass or volume.

put on zero

Active

24

W24 - Batch neg. flow
Check flow direction

Negative flow direction during batch.

Active

Active

30

W30 - Flowsaturated
Adjust max. flow

Flow is above Q

max.

settings.

Max. 120 %

Active

31

W31- Empty pipe

Pipe is empty.

Zero

Active

32

W32 - Temp. to high

The temperature of the fluid has exceeded the max.

Adjust temperature

temperature rating of the sensor (180

°C).

Active

Active

33

W33 - Temp. to low

The temperature of the fluid has exceeded the min.

Adjust temperature

temperature rating of the sensor (-50

°C).

Active

Active

34

W34 - Zero Adj. fail

Zero-point values are out of limits because of flow in the

Check flow = zero

sensor. Check zero flow conditions, valves, pumps etc.

Active

Active

35

W35 - Current Out 1

Current output exceeds 120%. Ensure that the sensor is

Check max. settings

correctly sized and check maximum flow setting.

Active

Active

36

W36 - Freq/Pulse Out1

Freq/Pulse output exceeds 120%. Ensure that the sensor

Check max. settings

is correctly sized and check maximum flow setting.

Active

Active

40

P40 - SENSORPROM

â

Insert

SENSORPROM

â

unit not installed.

Active

Active

41

P41 - Parameter range

A parameter is out of range.

Switch off and on

The error will disappear at the next power-on.

Active

Active

42

P42 - Current output

Current loop is disconnected or the loop resistance

Check cables

is too big.

Active

Active

43

P43 - Internal error

Internal error.

Switch off and on

Active

Active

49

P49 - Protec. viol.

Too many errors occured at the same time.

Switch off and on

Some errors are not detected correctly.

Active

Active

50

P50 - Temp. cable

Error in temperature sensor, check cables and

Check cable

connectors.

Active

Active

51

P51 - Pick-up 1

Pick-up 1 amplitude too low. Check cables or application

Check cable/install.

for damping (air/gas in liquid).

Active

Active

52

P52 - Pick-up 2

Pick-up 2 amplitude too low. Check cables or application

Check cable/install.

for damping (air/gas in liquid).

Active

Active

60

F60 - CAN comm. error

CAN bus communication error. An add-on module, the

Converter/add-on module

display module or the converter is defect

Zero

Inactive

61

F61 - SENSORPROM

â

err.

SENSORPROM

â

data is not reliable.

Replace

Active

Active

62

F62 - SENSORPROM

â

ID

SENSORPROM

â

unit ID doesn't match instrument ID.

Replace

The SENSORPROM

â

unit is from another instrument.

Zero

Inactive

63

F63 - SENSORPROM

â

It is not possible to read from the SENSORPROM

â

Replace

unit .

Active

Active

70

F70 - Pick-up phase

Check cables/polarity.

Active

Active

71

F71 - Driver phase

Check cables/polarity.

Active

Active

80-83

F80, 81, 82, 83 - Internal error Restart or replace.

Active

Active

84

F84 - Sensor level

Pick-up amplitude saturated.

Active

Active

97

F97 - Add-on module to old

Replace.

Active

Active

Error code level: W = Warning F = Fatal P = Permanent

ERROR

CODES