FLUID COMPONENTS, INTL

FlexCOR

TM

Model CMF Series Coriolis Flow Meter 4 Doc. No. 06EN003318 Rev. B

Locate the instrument as far away as possible from components that generate

mechanical vibration in the piping.

Vibration

Cross-Talk

If the instruments are located close to each other, e.g. in the same pipe section,
the instruments may interfere with each other's measurements, especially with

low flow applications. Install a flexible connection between instruments,

instead of a hard pipe connection.

Avoid mounting the instrument on the same steel frame. I.e. insulate the meters
mechanically.

CMF- B Through F SENSOR INSTALLATION

Horizontal Mounting in Pipe

Horizontal mounting
for Gas applications
Not for Liquid applications

Horizontal mounting
for Liquid applications

Direction of Flow

Static back pressure min. 0.1 bar

Valve

Reverse

Forward

Valve

Ok for gas
or liquid
applications

Ok for gas
applications
Only

Vertical Mounting in Pipe

Direction of

Vertical Flow

Direction of flow

Valve

Static back
pressure min.
0.1-0.2 bar

Locate the flowmeter as far away as possible from
components that generate mechanical vibration in the

piping.

Vibration

Ensure that there is no direct connection with them e.g.

by using flexible connections. The flowmeter can be
located after a bend.

Mounting the CMF-B through F sensors are very similar to mounting the CMF-A. A pictorial account of mounting will be given.