Doc. No. 06EN003318 Rev. B 7 FlexCOR

TM

Model CMF Series Coriolis Flow Meter

FLUID COMPONENTS, INTL

Zero-Point Adjustment for CMF-A Through F

Zeroing the instrument is normally not required in the field. However, if necessary the FlexCOR features an auto-zero setting.

Refer to Page 25 and 62 of the handbook document 06EN003327 for instructions on resetting the zero point in the field.

The keypad is used to set the flowmeter. The function of the keys are as follows:

TOP UP KEY

This key (hold 2 sec.) switches between operator
menu and setup menu. When In the converter

setup menu, a short press will cause a return to

the previous menu.

FORWARD KEY

This key steps forward through the menus. It is

the only key normally used by the operator.

BACKWARD KEY

This key is steps backward through the menus.

CHANGE KEY

This key changes settings or numerical values.

SELECT KEY

This key selects the numerals to be changed.

LOCK/UNLOCK KEY

This key allows the operator to change settings
and accesses submenus.

Keypad and Display Layout

Prom Insertion into a Wall Instrument The remote mounted transmitter requires the

installation of a

SENSORPROM

.

Open the transmitter as shown previously.
Remove the mother board by unscrewing a
captive screw found in the center of the board.
Pull up on the board.
Press the

SENSORPROM

onto the metal pronged

structure inside the enclosure. The

SENSORPROM

label must face the closest enclosure wall.
Press the mother board on top of the

SENSORPROM

and tighten the captive screw.

Close the transmitter.

Prom Insertion into a IP66 Mount Instrument

Open the transmitter as shown previously.
Install the

SENSORPROM

as shown, with the label facing the installer.

Close the transmitter.

SENSORPROM

TRANSMITTER MENU MANIPULATION

The display is alphanumeric and indicates flow values, flowmeter settings and error messages.
The upper line is for primary flow readings and will always show either mass flow rate, volume flow rate, density,
temperature, totalizer 1 or totalizer 2. The line is divided into 3 fields.
S: Sign field

P: Primary field for numerical value

U: Unit field
The center line is the title line (T) with individual information according to the selected operator or setup menu.