Introduction

The test methodology and equipment described herein can be used to gain some confidence in the
measurement accuracy of the MA24106A Power Sensor. This is accomplished by comparing the sensor to
another sensor with a specified cal factor and linearity performance or uncertainty. General commercially
available equipment is used for these tests; however, these procedures are not sufficiently accurate to verify
sensors to factory specification. Therefore, sensor test limits in these procedures are set appropriately to the
specified comparison equipment. All tests should be performed at an ambient temperature of 20ºC to 25ºC.

Refer to the following sections for required equipment and test procedures:

Required Equipment
VSWR Pretest
Calibration Factor Test
Linearity Test

5-2

Required Equipment

Note: Calibration and verification of high accuracy Power sensors requires substantial investment in both skill

and equipment. For calibration, calibration verification, and to maintain the factory specifications of your
power sensor, please send sensors to qualified Anritsu Customer Service Centers.

Table 5-1.

Required Equipment

Equipment Description

Manufacturer and

Model

Critical Specifications

Vector Network Analyzer (Pretest)

Anritsu MS4624B
or equivalent

Reflection Coefficient
Uncertainty

0.013 to 2 GHz

Uncertainty

0.020 to 6 GHz

Synthesizer
(Cal. Factor and Linearity Tests)

Anritsu MG3692
or equivalent

Output Power: >+20 dBm 0.05 to 6 GHz
Output Power Setting Resolution: 0.01 dBm
Harmonics:

–40 dBc

Source VSWR

2

Reference Power Meter
(Cal. Factor and Linearity Tests)

Anritsu ML2438
or equivalent

Instrumentation Accuracy

0.5%

Reference Power Sensor
(Cal. Factor and Linearity Tests)

Anritsu MA2422B
or equivalent

NIST Calibration or equivalent

10 dB K Attenuator
(Linearity Test)

Anritsu 41KC-10

VSWR

1.15 to 6 GHz

6 dB K Attenuator
(Cal. Factor Test)

Anritsu 41KC-6

VSWR

1.15 to 6 GHz

Adapter N(f) to K(f)
(Cal. Factor and Linearity Tests)

Anritsu 34ANF50 and
34AS50

VSWR

1.05 to 6 GHz

Power Splitter
(Linearity Tests)

Anritsu K241B

Effective Output VSWR <1.45 to 6 GHz

Power Sensor Under Test

MA24106A

See Datasheet, 11410-00424

Personal Computer

Any

See Chapter 2

Advertising