Run, save, and print reports – Luminex xPONENT 4.2 for FLEXMAP 3D User Manual

Page 13

Advertising
background image

4. Vortex the xMAP calibrator, verification, and fluidics containers for 30 seconds to ensure

homogeneity. Do not dilute xMAP calibrator, verification, or fluidics agents.

5. Click Eject on the status bar.

6. Place the off-plate reagent block on the plate holder. Load a strip well into the block.

7. Load wells with at least five drops of the Calibration, Verification, and Fluidics beads, as

indicated on the plate image and in the kit insert instructions.

NOTE: To ensure you get the necessary bead count, invert the calibrator

and verifier vials perpendicular to the plate as you add drops to the
wells. This ensures the maximum fluid drop size is dispensed into
the wells.

8. Click Retract.

9. Click Run.

Run, Save, and Print Reports

xPONENT

®

can provide information in six different report types:

• Batch

• Protocol

• Calibration and Verification

• Performance Verification

• System Log

• Advanced (User)
It can also format your batch or multibatch results in a variety of export formats.

9