Creating a kit – Luminex xPONENT 4.2 for FLEXMAP 3D User Manual

Page 17

Advertising
background image

11. Add standards, controls, unknowns, and maintenance steps to the plate layout.

12. Click Save to save the protocol.

Creating a Kit

To create a kit:

1. Open the Protocols page, then open the Protocols tab.

2. Select the protocol that you want to use for the kit, then click New Std/Ctrl. The Std/Ctrl

Details tab opens.

3. Type the name of the kit in the Name box, the lot number in the Std/Ctrl Kit Lot# box,

the expiration date using MM/DD/YYYY format in the Expiration box, and the
manufacturer in the Manufacturer box.

4. Click Apply Std Lot if you want to apply a standard lot. The Select Lot dialog box

opens. Click a lot and select OK.

5. Click Apply Ctrl Lot to apply a control lot. The Select Lot dialog box opens. Select a lot

and click OK.

6. Alternatively, type the appropriate information in the Assay Standard Information and

Assay Control Information sections. The number of standards, controls, or both in
these sections is defined in the protocol. If your batch uses controls, select Expected,
Low or High from the Show Value options. Use the Apply Values arrows to apply
values down or across the range of analytes.

7. Click Save.