Meriam M130 T_C Thermocouple Calibrator User Manual

Page 14

Advertising
background image

M130 – T/C Calibrator

Instruction Manual Z1017-IR

E.O. 6566 Sept. 2008

www.meriam.com

Page 14 of 21

Engineering Units Select

Engineering Units are conveniently selected when actively measuring or in
Manual Simulation mode. Press the W UNITS key to display any of the
following: ºC, °F, ºR, K. NOTE: to display mV units, select the mV option
under MEAS SETUP / TC TYPE. The W UNITS key will not change
display units for this TC TYPE.



Record Mode

The M130 can store up to 240 temperature measurements in a single record
(REC) session. For maximum flexibility, a REC session can be:

• Automatic - the current value is automatically stored every 5 seconds, for

up to 20 minutes.

• Manual - the current value is stored every time the Enter key is pressed, up

to 240 times.


Both types of REC sessions can store between 1 and 240 measurements. The
measurement data is preserved in non-volatile memory until another REC
session is started.


RECORD Data:

To start a manual record session from Measure Mode, press SETUP key and
then W or X key until RECORD is displayed. Press ENTER and then W or X
key until MANUAL is displayed. Press ENTER. Previously stored
information is deleted and M130 enters Record Manual Mode. The REC icon
flashes in upper right. Press Enter to save a displayed value to memory.
Pressing SETUP key ends the session.

To start an automatic record session from Measure Mode, press SETUP key
and then W or X key until RECORD is displayed. Press ENTER and then W
or X key until AUTO is displayed. Press ENTER. Previously stored
information is deleted and M130 enters Record Auto Mode. The REC icon
flashes in upper right. Values are stored every 5 seconds until the SETUP key
is pressed to end the session or 240 values are saved.

REC is displayed on the lower display line each time a sample is stored. After
240 measurements have been recorded or the SETUP key is pressed, the REC