Page 17 of 21


T/C Measurement & Simulation Accuracy Table*

Find T/C type in left column and follow the row across to the temperature range of
interest. Then follow the column up to determine the potential error in degrees C.

T/C

Type

±3.0°

C

±2.0°

C

±1.0°

C

±0.5°

C

±0.4°

C

±0.3°

C

±0.2°

C

±0.1°

C

E

-200 to
-179°C

-179 to

15°C

15 to

1000°C

J




-200 to
-165°C

-165 to

1200°C


K




-200 to

1372°C






N

-200 to

110°C

110 to

1300°C


T

-200 to
-168°C

-168 to

-86°C

-86 to

346°C

346 to
400°C

R

-50 to

27°C

27 to

736°C

736 to

1768°C

S

-50 to

25°C

25 to

1179°C

1179 to
1768°C

B

250 to
295°C

295 to
605°C

605 to

1301°C

1301 to
1820°C


*Notes: 1. Accuracy stated above applies to all T/C measurements and all manual

(non-ramp) T/C simulations.

2. Cold junction effect at 25º C included in above table.